8712ET/ES and 8714ET/ES Performance Checks
RF Network Analyzers Performance Test Software
Service Guide 1-15
Getting Started
CAUTION Make a copy of the performance test software disk for day-to-day use. The
master copy should remain stored in a safe place. This will decrease the
likelihood of damage to the master disk.
NOTE The DOS version, which is used in conjunction with HP BASIC for Windows,
requires the proper setup of the GPIB interface card and drivers. This
document assumes that users are familiar with HP BASIC for Windows, that
it is running, and that it has been properly installed and configured.
First, find all serial numbers of the calibration devices required, such as external step
attenuators, sensors, and fixed attenuators for both 50 and 75 devices (if applicable).
Verify the GPIB addresses you will be using, as well as the mass storage directories and
locations.
If you plan to operate this program only from the supplied floppy disk, insert the disk into
the computer's disk drive and skip step 1 next.
Performance Test Software Installation
1. To install this program to the hard drive, either copy all files (except “INSTALL”) to the
hard drive, or load and run the “INSTALL” program. The INSTALL program will
automatically copy all required files to your specified destination drive or directory
(HFS, SRM or LIF format). The INSTALL program is only contained in the LIF
software version.
NOTE Any directory name used must be less than 80 characters long, and preferably
less than 60 characters, for best visibility.
2. Set the computer's “Mass Storage Is” (MSI) command to the directory or drive location
of the main program file (CAL871XE).
3. GET and RUN the CAL871XE program. The program will display the initial screen
with the software part number, revision date and current time/date. If the current time
or date is incorrect, select the softkey to set it.
4. Press . Remember that the key can also be used as softkey #1.
5. Press to enter the configuration screen.
This menu allows you to enter all the specific information that pertains to your
particular configuration (such as test equipment, addresses, and serial numbers). The
configuration file's selection descriptions and their initial default values are listed
below.
SET TIME
RESUME
Return
CONFIG
Vedere la pagina 22