8712ET/ES and 8714ET/ES Performance Checks
RF Network Analyzers Performance Test Software
Service Guide 1-29
PT06: Absolute Power Accuracy
This test verifies the accuracy of the B* broadband detector from the maximum input level
of +16dBm down to the specified noise floor. The test is designed to be performed at
30MHz, however 15 MHz may be used if the frequency range of the signal source does not
extend to 30 MHz. The 8116A is the preferred signal source, however the 33120A may be
used as an alternative. An external 110 dB step attenuator is used to vary the test signal
over the power range to be tested.
Test Process.Connect the step attenuator to the output of the signal generator. Set the
frequency to 30 MHz (preferred) or 15 MHz. Connect a 6 dB pad to the output of the step
attenuator. For 75 analyzers, use a minimum loss pad in place of the 6 dB attenuator.
Calibrate the power meter and use it to measure the output from the 6 dB pad. Set the
step attenuator to 0 dB. Adjust the output from the signal generator so that the power
output from the 6 dB pad is +16.0 dBm. Record the exact reading. Use calibration data
from the power sensor to compensate for the frequency difference between the power
meter calibration and the signal generator. Disconnect the power sensor and connect the
6dB pad to the RF IN/PORT 2 test port on the analyzer. Set the system bandwidth to
narrow (250 Hz). Repeat the previously used steps for the signal generator output, and
record the readings on B*. Switch the step attenuator as necessary to provide levels from
+16 dBm to44 dBm to the analyzer. Record the B* reading for each level and correct the
readings using the calibration data for the step attenuator. Averaging may be used as
needed. Each power reading on B* must meet the specified accuracy until the noise floor is
reached.
Specification Verified.This test verifies the “Test Port Input: Power Accuracy”
characteristic inTable1-9 starting on page1-63. In some cases, it may be performed at 15
MHz instead of the specified 30 MHz.
If This Test Fails.If you are using a programmable attenuator, first verify that the
correct 40 dB section of the step attenuator is being used. See“The step attenuator's 40 dB
card selection (3)” on page 1-17. If the correct attenuator is being used, perform the“B*
Amplitude Correction, Adjustment #110” on page 2-21.