Performance Checks 8712ET/ES and 8714ET/ES
System Performance RF Network Analyzers
1-42 Service Guide
Table 1-3 System Performance, Corrected, 75, 2-Port Calibration (continued)
1. These specifications apply for measurements made using “fine”
bandwidth, no averaging, and at an environmental temperature
of 25°±5 °C, with less than 1 °C deviation from the calibration
temperature.
2. Assumes using an 85039B cal kit, and a DUT with a center pin
conforming to the 0.77 to 0.86 mm limits.
3. For transmission measurements, the effect of crosstalk is
neglected and S12 = S21 for S21 <1.0, S12 = 1 / S21 for S21 > 1.0.
8712ES/8714ES (75 )
85039B (Type-F, 75 ) Cal Kit, User 2-Port Calibration
Description
Specification
1
(in dB)
300 kHz
to 1.0 GHz
1.0 GHz
to 3 GHz
Directivity 40 40
Source Match 30 30
Load Match 40 40
Reflection Tracking
2
±0.071 ±0.025
Transmission Tracking
2
±0.073 ±0.090
Transmission Uncertainty (Specification)
1,3
Reflection Uncertainty (Specification)
1