Performance Checks 8712ET/ES and 8714ET/ES
Test Port Output RF Network Analyzers
1-60 Service Guide
Table 1-8 Test Port Output (continued)
1. Maximum power setting is the highest power setting allowed that maintains the
network analyzer’s specified level accuracy.
2. Option 1E1 adds a 60 dB step attenuator to the 8712ET/8714ET.
8712ET/ES and 8714ET/ES Test Port Output
Description Specification
Supplemental
Information
Output Power (continued)
<1 GHz >1 GHz
Maximum Power Setting:
1
8712ET:
50 :
With Attenuator Option 1E1
2
15 dBm 12 dBm
Without Attenuator Option 1E1 16 dBm 13 dBm
75 :
With Attenuator Option 1E1
2
12 dBm 9 dBm
Without Attenuator Option 1E1 13 dBm 10 dBm
8712ES:
50 13 dBm 10 dBm
75 10 dBm 7 dBm
8714ET:
50 :
With Attenuator Option 1E1
2
10 dBm 9 dBm
Without Attenuator Option 1E1 11 dBm 10 dBm
75 :
With Attenuator Option 1E1
2
7 dBm 6 dBm at <2 GHz 6 dBm at >2 GHz, char.
Without Attenuator Option 1E1 8 dBm 7 dBm at <2 GHz 7 dBm at >2 GHz, char.
8714ES:
50 9 dBm 7 dBm
75 6 dBm 4 dBm at <2 GHz 4 dBm at >2 GHz, char.