Agilent Technologies 85032B/E 50 Specifiche Pagina 96

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
Vedere la pagina 95
8712ET/ES and 8714ET/ES Adjustments
RF Network Analyzers Introduction
Service Guide 2-3
Required Equipment
Table 2-1 Required Equipment for Adjustments
1. An 85032B Option 001 is a subset of the standard 85032B. This option eliminates four 7-mm
adapters that are not needed, resulting in a significant cost savings.
2. The required attenuators depend on the firmware revision of the analyzer. The attenuators
listed above apply to all versions of firmware.
3. The frequency counter is useful when very accurate frequency measurements are needed. How-
ever, if you are using a synthesized spectrum analyzer, the frequency can be measured by using
the spectrum analyzer. Note that the frequency measurement using a spectrum analyzer will
not be quite as accurate as a frequency counter, but is sufficient in most cases. (An 8560 series
spectrum analyzer may be used.)
Description HP/Agilent Model #
50
STD
75 Notes
Power meter
437B, 438A
E4418A, E4419A
✓✓
Do not use an 436A.
Power sensor, 50 8482A
✓✓
Low power sensor
8481D or 8484A
✓✓
For the ES models and
ET Option 1E1.
Cal kit type-N 50
85032B Opt 001
1
Cal kit type-N 75
85036B
6 dB attenuator 50 8491A/B/C Opt 006
Required for adjustments only if
analyzer does not have an internal
attenuator installed.
10 dB attenuator 50
2
8491A/B/C Opt 010
Required for adjustments only if
analyzer does not have an internal
attenuator installed.
10 dB attenuator 75
2
0955-0767
Required for adjustments only if
analyzer does not have an internal
attenuator installed. (May substitute
2 min loss pads.)
20 dB attenuator 50 8491A/B/C Opt 020
Required for adjustments only if
analyzer does not have an internal
attenuator installed.
20 dB attenuator 75
2
0955-0768
Required for adjustments only if
analyzer does not have an internal
attenuator installed. (May substitute
2 min loss pads and a 50 10 dB pad.)
Minimum loss pad 11852B
Frequency counter 535xB or equivalent
——
May not be required.
3
Precision cable 50 8120-8862 or 8120-6469
Precision cable 75 8120-8898 or 8120-6468
Voltage reference source 08712-60031
✓✓
Included in the service kit, part
number 08712-60012.
10 dB attenuator 50 8491A, Option 010
Generally not needed; occasionally
required for adjustment #104.
BNC cable, 50 8120-1839 or equivalent
✓✓
Adapter, type-N (m) to
BNC (f)
1250-0780 or equivalent
✓✓
Vedere la pagina 95
1 2 ... 91 92 93 94 95 96 97 98 99 100 101 ... 268 269

Commenti su questo manuale

Nessun commento