8712ET/ES and 8714ET/ES Performance Checks
RF Network Analyzers System Performance
Service Guide 1-45
Table 1-4 System Performance, Corrected, 50, T/R Calibration (continued)
1. Option 1E1 adds a 60 dB step attenuator to the 8712ET
and 871414ET.
2. These specifications apply for measurements made
using “fine” bandwidth, no averaging, and at an envi-
ronmental temperature of 25°±5 °C, with less than 1 °C
deviation from the calibration temperature.
8712ET/8714ET (50 ) with Attenuator Option 1E1
1
85032B/E (Type-N, 50 ) Cal Kit, T/R Calibration
Description
Specification
2
(in dB)
300 kHz
to 1.3 GHz
1.3 GHz
to 3 GHz
Directivity 50 47
Source Match:
Reflection (One-Port Cal) 42 36
Transmission (Enhanced Response Cal) 42 36
Transmission (Response Cal) 21 15
Load Match 18 15
Reflection Tracking ±0.02 ±0.02
Transmission Tracking:
Enhanced Response Cal ±0.039 ±0.055
Response Cal ±0.13 ±0.3
Vedere la pagina 52