8712ET/ES and 8714ET/ES Service Related Menus
RF Network Analyzers GPIB Command Reference for Service
Service Guide 5-25
Table 5-6 Alphabetical SCPI Command Summary
Command Form Description
DIAGnostic:BANDcross
[:STATe]<OFF|0|ON|1>
NR1 (8714ET/ES only) Turns
bandswitching on or off.
DIAGnostic:CCONstants:COPY Command Copies correction constants from
EPROM to RAM.
DIAGnostic:CCONstants:DEFaults Command Restores the default correction
constants to EPROM and RAM.
DIAGnostic:CCONstants:ERASe: EEPRom Command Erases EPROM.
DIAGnostic:CCONstants:LOAD Command Loads correction constants from
internal disk drive.
DIAGnostic:CCONstants:INSTalled? Query only
NR1
Queries if correction constants are
installed. Returns a “1” if true,
and a “0” if false.
DIAGnostic:CCONstants:STORe:DISK Command Stores correction constant data to
a file on the internal disk drive.
DIAGnostic:CCONstants:STORe:EEPRom Command Stores correction constant data
from the RAM buffer into flash
EPROM.
DIAGnostic:CPU:MEMory[:WORD]
<addr num><data num>
NR1 Writes a 16-bit data value to a
68030 address. Queries returns an
integer with the 16-bit value at
the selected address.
DIAGnostic:SNUMber <string> STRING Stores a serial number into
memory. The string must be 10
characters long. The query form
returns the current serial number.
*IDN? Query
STRING
Returns a string that uniquely
identifies the analyzer, including
model number, serial number, and
firmware revision.
*OPT? Query
STRING
Returns a string identifying the
installed options, in the form
“1E1, 1C2, etc.…”
SENSe[1|2]:CORRection[:STATe]
<ON |1|OFF|0>
NR1 Switches measurement
calibration ON or OFF.
TEST:RESult? Query
CHAR
Queries the result of the selected
adjustment or self-test. Response
will be NULL|PASS|FAIL.
TEST:SELect <num> NR1 Selects the adjustment or self-test
to execute.