Performance Checks 8712ET/ES and 8714ET/ES
System Performance RF Network Analyzers
1-48 Service Guide
Table 1-4 System Performance, Corrected, 50, T/R Calibration (continued)
1. These specifications apply for measurements made using “fine”
bandwidth, no averaging, and at an environmental temperature
of 25°±5 °C, with less than 1 °C deviation from the calibration
temperature.
2. For transmission measurements, the effect of crosstalk is
neglected and S12 = S21 for S21 <1.0, S12 = 1 / S21 for S21 > 1.0.
8712ET/8714ET (50 ) without Attenuator
85032B/E (Type-N, 50 ) Cal Kit, T/R Calibration
Transmission Uncertainty: Enhanced Response Calibration
(Specification)
1,2
Transmission Uncertainty: Response Calibration (Specification)
1,2
Reflection Uncertainty: One-Port Calibration (Specification)
1