8712ET/ES and 8714ET/ES Service Related Menus
RF Network Analyzers Meas Cal Options Menu
Service Guide 5-15
Meas Cal Options Menu
The analyzer uses arrays of calibration data to improve the measurements made by the
hardware. This menu allows you to disable the calibration, and view the calibration data
arrays.
: Allows the cal arrays to be used in processing measurement data.
This function controls both the user-performed calibrations and the default factory cal.
This includes transmission, reflection, and frequency response cal arrays (both broadband
and narrowband). This function provides a way to examine the uncorrected performance of
the hardware in the analyzer. When this function is off, “Cx” is displayed near the upper
right-hand corner of the display.
: Allows you to switch off the power cal for both the internal and
external broadband detectors.
: Selects the master (uninterpolated) cal arrays from the master to
be displayed. These are the cal arrays from the currently active calibration (default, full
band, or user-defined).
: Selects the interpolated cal arrays from the master to be
displayed. These are the cal arrays from the currently active calibration (default, full band,
or user-defined). If the frequency range of the analyzer has not been changed since the last
calibration was performed, then this is the same as the master.
: Selects the scalar cal arrays from the master to be displayed.