8712ET/ES and 8714ET/ES Performance Checks
RF Network Analyzers System Performance
Service Guide 1-39
Table 1-2 System Performance, Corrected, 50, 2-Port Calibration (continued)
1. These specifications apply for measurements made using “fine”
bandwidth, no averaging, and at an environmental temperature of
25°±5 °C, with less than 1 °C deviation from the calibration tem-
perature.
2. For transmission measurements, the effect of crosstalk is
neglected and S12 = S21 for S21 <1.0, S12 = 1 / S21 for S21 > 1.0.
8712ES/8714ES (50 )
85033D (3.5 mm, 50 ) Cal Kit, User 2-Port Calibration
Description
Specification
1
(in dB)
300 kHz
to 1.3 GHz
1.3 GHz
to 3 GHz
Directivity 46 43
Source Match 44 41
Load Match 46 43
Reflection Tracking ±0.016 ±0.008
Transmission Tracking ±0.04 ±0.05
Transmission Uncertainty (Specification)
1,2
Reflection Uncertainty (Specification)
1