8712ET/ES and 8714ET/ES Performance Checks
RF Network Analyzers System Performance
Service Guide 1-51
Table 1-5 System Performance, Corrected, 75Ω, T/R Calibration (continued)
1. Option 1E1 adds a 60 dB step attenuator to the
8712ET/8714ET.
2. These specifications apply for measurements made
using “fine” bandwidth, no averaging, and at an envi-
ronmental temperature of 25°±5 °C, with less than 1 °C
deviation from the calibration temperature.
8712ET/8714ET (75 Ω) with Attenuator Option 1E1
1
85036B/E (Type-N, 75 Ω) Cal Kit, T/R Calibration
Description
Specification
2
(in dB)
300 kHz
to 1.3 GHz
1.3 GHz
to 3 GHz
Directivity 48 43
Source Match:
Reflection (One-Port Cal) 41 35
Transmission (Enhanced Response Cal) 41 35
Transmission (Response Cal) 21 15
Load Match 18 15
Reflection Tracking ±0.021 ±0.02
Transmission Tracking:
Enhanced Response Cal ±0.04 ±0.062
Response Cal ±0.125 ±0.295