8712ET/ES and 8714ET/ES Assembly Replacement
RF Network Analyzers A2 CPU Board Assembly Replacement
Service Guide 3-9
NOTE In the unlikely event that the bootROM has failed, then the replacement
board (which has the transferred bootROM installed) will also “fail.” A
“minimum configuration” bootROM is included in the service kit to help
isolate this type of failure. If it is proven that the bootROM has failed, contact
your nearest Agilent Technologies sales or service office for assistance.
NOTE If you have ordered a rebuilt-exchange assembly, the failed CPU assembly
must be returned with the SIMM module installed.
Installing the CPU Board Assembly
To install the CPU board assembly into the analyzer, reverse the steps of the “Removing
the CPU Board Assembly” procedure.