Assembly Replacement 8712ET/ES and 8714ET/ES
Introduction RF Network Analyzers
3-2 Service Guide
Introduction
This chapter describes how to disassemble and reassemble your analyzer for the purpose
of replacing a failed assembly. It also contains specific instructions for replacing each
assembly.
A section on ordering parts, “How to Order Parts” on page 3-24, is included at the end of
this chapter.
NOTE Some individual components of each assembly may also be replaced. See
Chapter 7, “Parts List,” for a list of the parts that are available for
replacement.
WARNING Parts of this instrument may have sharp edges. Work carefully to
avoid injury.
WARNING These servicing instructions are for use by qualified personnel only.
To avoid electrical shock, do not perform any servicing unless you
are qualified to do so.
Required Tools
TORX drivers, sizes T-10 and T-15
5/8-inch wrench (for BNC nuts)
5/16-inch wrench (preferably the modified wrench included in the service kit)
5/16-inch nut driver (required only to replace the display)
5/16 inch 10 in-lb torque wrench
7/32 inch wrench (included in service kit)