Agilent Technologies 85032B/E 50 Specifiche Pagina 36

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
Vedere la pagina 35
Performance Checks 8712ET/ES and 8714ET/ES
Performance Test Software RF Network Analyzers
1-28 Service Guide
PT05: Power Flatness
This test checks the maximum variation in output power across the frequency range of the
analyzer.
Test Process.Connect a calibrated power sensor to the RF OUT/PORT 1 test port.
Determine the power levels to test:
1.Refer to “Test Port Output: Power Sweep Ranges” inTable1-8 starting on page1-57.
2.Find the entry for the model and option configuration of the analyzer under test.
3.In the Attn = 0 dB” column of the table, the power sweep range is defined in the form
ntoPmax”. Usen as one power level to measure, use Pmax as another power level to
measure, and use the midpoint between the two as the third power level to measure.
4.If the analyzer under test is an ET model with Option 1E1, or an ES model, then also
use a midpoint power level from each of the following three columns: Attn = 10 dB”,
Attn = 20 dB” and Attn = 30 dB”.
5.For each separate power level, record the reading on the power meter at 27 different
frequencies, across the frequency range of the analyzer. Correct the power readings
using the frequency flatness data for the power sensor. The difference between the max
and min values for the 27 readings should be within the specified flatness limit as
stated in “Test Port Output: Level Accuracy inTable1-8 starting on page1-57. Note
that the specifications are listed in the form of±n dB. This equates to an allowable 2n
variation (maxmin). Measurement of power levels below30 dBm are outside the
specified range of the power sensor and should not be included in the test.
6.For ES models, make the same set of measurements in the reverse transmission
direction (S12) with the power sensor connected to PORT 2.
NOTE Be sure to zero the power meter before making any series of measurements
below19 dBm.
A second test is performed by the test software to check for obvious power level
inconsistencies that might not be detected by the specification check. It is a precaution to
ensure that no power holes (drop-outs) exist. A through cable is connected and the
analyzer is swept from 10 MHz to its high-end frequency using 1601 points per sweep. The
resulting B* trace is checked for any significant power holes that could have been missed
during the previous section of this test. For all ES models and for ET models with
Option1E1, repeat the full-sweep test for the three power levels tested in step 4 above.
When power holes are detected by this test, it is considered to be a failure of the
specification.
Specification Verified.This test verifies the “Test Port Output: Level Accuracy”
specification inTable1-8 starting on page1-57. Notice that values for 75 instruments
above 2GHz are nonwarranted characteristics,not warranted specifications.
If This Test Fails.Repeat the test, verifying that all connections are secure. If the test
still fails, perform the“Source Power Correction, Adjustment #104” on page 2-15.
Vedere la pagina 35
1 2 ... 31 32 33 34 35 36 37 38 39 40 41 ... 268 269

Commenti su questo manuale

Nessun commento