8712ET/ES and 8714ET/ES Performance Checks
RF Network Analyzers Test Port Input
Service Guide 1-69
Table 1-9 Test Port Input (continued)
1. Trace noise is defined for a transmission measurement in CW
mode, using a “through” cable having 0 dB loss, with the source
set to 0 dBm, and the analyzer’s averaging function toggled off.
8712ET/ES and 8714ET/ES Test Port Input
Description Spec. Typical
System Bandwidths:
Narrow
(250 Hz)
Narrow
(250 Hz)
Med Wide
(4000 Hz)
Wide
(6500 Hz)
Trace Noise
1
8712ET/8712ES:
Narrowband:
Magnitude 0.01 dB rms 0.03 dB-pp 0.12 dB-pp 0.28 dB-pp
Phase 0.2 Deg-pp 2.5 Deg-pp 3.4 Deg-pp
Broadband:
Magnitude 0.01 dB rms 0.01 dB-pp 0.02 dB-pp 0.15 dB-pp
8714ET/8714ES:
Narrowband:
Magnitude 0.01 dB rms 0.02 dB-pp 0.06 dB-pp 0.23 dB-pp
Phase 0.2 Deg-pp 0.8 Deg-pp 1.8 Deg-pp
Broadband:
Magnitude 0.01 dB rms 0.01 dB-pp 0.03 dB-pp 0.16 dB-pp