Adjustments 8712ET/ES and 8714ET/ES
B* Amplitude Correction, Adjustment #110 RF Network Analyzers
2-22 Service Guide
Procedure
1. Press
.
2. This is a fully automated adjustment handled by the internal firmware. Follow the
instructions on the analyzer's display.
The analyzer presets after the test is done.
3. If you are not going to make any more adjustments, press
.
a. Insert a formatted disk into the internal disk drive.
b. Press .
This creates a file (or writes over an existing file) where the correction constants are
stored.
PRESET SYSTEM OPTIONS
110 ENTER
SYSTEM OPTIONS