Table 1-5 System Performance, Corrected, 75, T/R Calibration (continued)
1. These specifications apply for measurements made using
“fine” bandwidth, no averaging, and at an environmental
temperature of 25°±5 °C, with less than 1 °C deviation from
the calibration temperature.
8712ET/8714ET (75 ) without Attenuator
85036B/E (Type-N, 75 ) Cal Kit, T/R Calibration
Description
Specification
1
(in dB)
300 kHz
to 1.3 GHz
1.3 GHz
to 3 GHz
Directivity 48 43
Source Match:
Reflection (One-Port Cal) 41 35
Transmission (Enhanced Response Cal) 41 35
Transmission (Response Cal) 23 19
Load Match 18 15
Reflection Tracking ±0.021 ±0.02
Transmission Tracking:
Enhanced Response Cal ±0.04 ±0.058
Response Cal ±0.11 ±0.2