Agilent Technologies 85032B/E 50 Specifiche Pagina 95

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
Vedere la pagina 94
Adjustments 8712ET/ES and 8714ET/ES
Introduction RF Network Analyzers
2-2 Service Guide
Introduction
This chapter contains procedures to adjust the analyzer. Generally these adjustment
procedures are only used when hardware has been replaced in the instrument, or if there
is a failure in the operators check or the performance tests (seeChapter 1). Refer to
Table3-1, “Post Repair Procedures, on page3-23 to determine which adjustment
procedures you need to do when a particular part has been replaced.
Adjustments #102 through #111 affect the correction constants that are stored in the
non-volatile memory of the instrument. As adjustments are performed, new correction
constants are created.The user controls when the new correction constants are stored in
the non-volatile memory. The frequency accuracy adjustment, which could require a
potentiometer adjustment, and the serial number (Adjustment #100) do not affect the
correction constants.
Before performing adjustments, turn on the analyzer and allow it to warm up for one hour.
Most adjustments are accessed through the analyzer's service menus.
You should perform the adjustment procedures in the order given here:
1.”Frequency Accuracy Adjustment”on page 2-5
2.”Switched Gain Correction, Adjustment #102”on page 2-10
3.”External Detector Gain Correction, Adjustment #103”on page 2-12
4.”Auxiliary Input Correction, Adjustment #111”on page 2-14
5.”Source Power Correction, Adjustment #104”on page 2-15
6.”B Amplitude Correction, Adjustment #105”on page 2-18
7.”Reflection (One-Port) Correction, Adjustment #107”on page 2-19
8.”Transmission Correction, Adjustment #106”on page 2-20
9.”B* Amplitude Correction, Adjustment #110”on page 2-21
10.”R* Amplitude Correction, Adjustment #108”on page 2-23
11.”R* Frequency Response Correction, Adjustment #109”on page 2-25
CAUTION Procedures 2 through 11 all generate correction constants. For efficient operation,
these correction constants must be handled properly. Please read the“Correction
Constants (CCs) Storage” on page2-4 carefully.
CAUTION The preset power level is user-definable. After pressing in any
adjustment procedure, change the power level to 0 dBm if necessary.
NOTE During the adjustments, the analyzer will display a list of instructions in which the
names for the connectors may vary. For user clarification, “REFLECTION,” “RF
OUT,” and “PORT 1” all refer to the connector closest to the analyzers display, while
“TRANSMISSION,” “RF OUT,” and “PORT 2” all refer to the connector farthest from
the display.
PRESET
Vedere la pagina 94
1 2 ... 90 91 92 93 94 95 96 97 98 99 100 ... 268 269

Commenti su questo manuale

Nessun commento