When loading a Result File for calculating Sigma vs. Tau,
the carrier frequency used for the measurement must be
entered in the Instrument Parameters display in order to
compute accurate values.
Detector/Discr. Input Frequency is the frequency of the input signal
connected to the HP 11848A's phase detector. The Detector/Discriminator
Input Frequency is used to:
Select the appropriate Low Pass Filter (LPF) in the HP 11848A Phase
Noise Interface. For frequencies < 95 MHz, the system enables the
2 MHz LPF to eliminate feedthrough of the fundamental signal and
the additive products out of the phase detector. When the 2 MHz LPF
is enabled, the maximum offset frequency the system can measure is
2 MHz.
Set the frequency of the Reference Source when the Reference Source
is under system control.
NOTE
When using an HP 11729C under System Control, the
HP 3048A will derive the correct Detector/Discr. Input
Frequency and enter it into the Instrument Parameter dis-
play. If the currently entered Detector/Discr. Input Fre-
quency is less than 95 MHz, and you are allowing the
HP 3048A to input the correct Detector/Discr. Input Fre-
quency by computing the value from a System Controlled
HP 11729C, then the measured offset range will be limited
to 2 MHz. To eliminate this problem, set the Detect/Discr.
Input Frequency to a frequency greater than 95 MHz.
2-32 instr. Params: $N w/o PLL