Description
To configure an RF Analyzer into the system, enter the Name (RF
Analyzer), Model Number, and HP-IB Address of the RF Analyzer into
the System Configuration Table. The serial number and option numbers
may also be entered for the RF Analyzer.
Connect the 50O Input of the RF Analyzer to the "Spectrum Analyzer"
PHASE DETECTOR OUTPUT port of the HP 11848A Phase Noise
Interface.
NOTE
If an RF Analyzer is not connected to the HP 11848A, the
Spectrum Analyzer output must be terminated in 50Q .
The system generates a 100 kHz Calibration Tone to calibrate the RF
Analyzer to match the HP 3561A prior to a new measurement. This tone
is used for all of the supported analyzers except the HP 3585A/B. (The
HP 3585A/B accuracy is comparable to that of the HP 3561A.) An error
message is displayed if the calibration factor measured during RF Analyzer
Calibration is > 5 dB.
The calibration is performed when the system is first "booted-up" and run
with the RF Analyzer in the System Configuration menu. If you change RF
Analyzers, use the BASIC command "SCRATCH C* to clear the present
calibration factor. Perform the New Measurement function to recalibrate
the new RF Analyzer.
rev.05SEP89
RF Analyzers 6-24.1