Table 8-3. Reference Messages (3 of 7)
6
7
8
9
Remove the calibration spur. If the spur signal is not removed, it will appear in the
measurement and could affect the measurement results.
Action:
Disconnect or turn off the spur source without disturbing the frequency or
amplitude settings of the source.
NOTE
// you wish to determine the frequency and amplitude of you spur source,
leave the source connected and turned on during the measurement. The
spur signal should appear in your measurement results. Note, however, that
a large spur signal can prevent the system from having the full dynamic
range necessary to make a valid noise measurement.
Offset the source and reference signals by at least 1 Hz and not more than
20 MHz. The system requires that two different signals be applied to the input ports
of the HP 11848A's Phase Detector. The frequencies of the two sources must differ as
follows:
With RF Spectrum Analyzer in System Offset Range: 1 Hz to 20 MHz
Without RF Spectrum Analyzer in System Offset Range: 1 Hz to 50 kHz
Action:
Adjust the frequency of the device-under-test (DUT) to offset it from the
reference source frequency.
If the DUT cannot be adjusted to create a frequency offset that is within the offset
range specified above, substitute a different source for the DUT. Adjust the substitute
source for the same power level as the DUT source it is replacing. Terminate any ports
in 50O that are not connected during the substitution. Reconnect the DUT source after
the calibration is completed.
Remove the beatnote and connect the system for noise measurement.
Action:
Readjust the device-under-test (DUT) to the frequency of the reference source
or reconnect DUT.
Phase lock has been lost. This message can be caused by any one of the following
conditions.
1.
Frequency drift > 25% of the Peak Tuning Range.
2.
A phase transient > 0.25 radians.
3. Out-of-Lock detector set by search oscillator.
Reference Messages 8-23