Scale Data
"Scale data to new carrier freq defined as: (f
c
)
*
[1]" (where f
c
is the defined
carrier freqeuncy in the Instrument Parameters menu), allows you to plot
results based on frequency changes in the carrier. This feature allows you
to plot the noise of the measured source to a translated frequency. The
carrier freqeuncy is shown on the same line opposite the field for entering
the scale factor. The scale factor is a direct multiple. Entering a factor-of-2
doubles the frequency, a factor-of-3 triples the frequency, and so on.
Plots made after the scale factor has been entered indicate noise for the
scaled freqeuncy. Noise levels increase for scale factors > 1 and decrease
for factors < 1.
rev.05SEP89
Scale Data 4-33
Vedere la pagina 255