How to Manage the Test Files
Storing a Test File
It is recommended that you complete the measurement before storing the
Test File to ensure that all of the entered parameters in the Measurement
Definition are correct. Also, when the Test File is stored after the
measurement has been completed, the correct calibration constants for
the measurement are stored as well. You can then perform a Repeat
Measurement directly after loading a Test File for measuring devices
having similar characteristics. The following steps describe the storing
process.
1.
Press the
Test Files
softkey to access the Test File Directory display.
2.
Move the cursor to the bottom of the File Name column. (If the page
is full, press Next Page to access a page with available entry space.)
NOTE
It is possible to lose data when using a hard disc or
SRM Mass Storage Device for storing the Test Files. If you
attempt to store more than 99 Test Files on one media, the
HP 3048A will destroy previously stored Test Files.
Disc storage is usually limited to approximately 50 Test
Files per disc.
3.
Enter a name for the hie. The HP 3048A will allow 47 characters in
the File Name column.
4.
Press the Store File softkey. The Test File will be stored in the Mass
Storage Media location defined for "Test Parameter Files'
Test Files 2-125