Agilent Technologies HP 8642A/B Specifiche Pagina 63

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 453
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 62
Table 1-2. HP 3048A System Option Specifications (3 of 9)
Spurious Signals
Spurious nonhar-
monicafly related
1
Subharmonically re-
lated (1/2, 3f/2, etc.)
Power line (60 Hz)
related to microphon-
ically generated
(within 300 Hz)
3
Harmonics
Frequency Range (MHz)
0.1 to
120
-90
dBc
none
-90
dBc
120 to
160
-100
dBc
none
-85
dBc
160 to
320
-96
dBc
none
-80
dBc
320 to
640
-90
dBc
none
-75
dBc
640 to
1280
-84
dBc
-70
2
dBc
-70
dBc
1280 to
2560
-78
dBc
-40
dBc
-65
dBc
<30 dBc
1
In the remote mode it is possible to have microprocessor clock-related spurious
signals spaced 3 MHz apart at an absolute level ot typically less than -145 dBm.
2
3/2 spurs not specified for HP 8662A carrier frequencies above 850 MHz.
^ At a 50 Hz line frequency, powerline or microphonicaily related spurious signals may
be up to 3 dB higher and appear at offsets as high as 1 kHz from the carrier.
Amplitude
Typical Maximum Output Level: +16 dBm
Modulation
Modulation Types: FM, AM (Pulse with the HP 8663A)
FM Deviation
Center Frequency (MHz)
0.1 -120
120-160
160 - 320
320 - 640
640-1280
1280-2560*
Maximum Peak Deviation dc FM (kHz)
100
25
50
100
200
400
'HP B663A Option 003 only
Indicated FM Accuracy: ±8% (7% for HP 8663A) of reading
plus 10 Hz (50 Hz to 20 kHz rates).
Input Impedance: HP 8662A: lkfi nominal, HP 8663A: 600fi.
Maximum Input Level: IV peak for specified accuracy.
Temperature Range: +15° to 35°C.
1-18 General Information
Vedere la pagina 62
1 2 ... 58 59 60 61 62 63 64 65 66 67 68 ... 452 453

Commenti su questo manuale

Nessun commento