You can set all fields on the RF segment table, but you must also be able to
set the values on the particular RF Analyzer you are using. The accuracy
of the measurement is dependent on the parameters in the RF Segment
Table. The RF Segment Table is stored with a stored Test File.
Table 6-8. RF Analyzer Default Segment Table
Plotted
Range (Hz)
100+3 - 500E+3
500+3 - 5E+6
5E+6 - 40E+6
0
0
0
Measurement
Range (Hz)
500E+3
5E+6
40E+6
Min.
Av'g
i
i
1
BW (Hz)
3.E+3
10.E+3
30.E+3
Video
BW (Hz)
10
30
100
Sweep
Time
20
20
20
Plotted Range
The Plotted Range defines which data will be plotted on a graph. It can
be the entire Measurement Range or a subset of the Measurement Range.
The Plotted Range is usually selected so that all the segments make a
continuous plot of the measured data.
Measurement Range
The Measurement Range is the range over which the analyzer will take
the measured data. The Measurement Range frequency must be less than
or equal to 40 MHz for all segments specified.
Bandwidth
The Bandwidth is the resolution bandwidth used for a particular segment.
The specified bandwidths should be at least as small as those specified in
the default segment table, but can be of finer resolution if required. User
RF bandwidths should be specified as 1 kHz, 3 kHz, 10 kHz, 30 kHz,
etc.
These bandwidths are available on all supported RF analyzers. If your
6-26 RF Analyzers
rev.05SEP89