Frequency Parameter Considerations
If the Carrier Frequency denned for the measurement is incorrect, the
Computed Output calculations may be in error. Verify that the Carrier
Frequency displayed on the Result Graph is the correct frequency.
Also,
when using an HP 11729C under system control, an incorrect
Carrier Frequency entry can result in an incorrect filter selection.
The HP 3048A will change the Detector/Discr. Input Frequency and
update the Instrument Parameter display when a Source is under
System Control and a frequency change was required to zero the
beatnote. If the change in frequency changes other parameters such
as Peak Deviation for DC FM, the HP 3048A will inform you of the
change and allow you to proceed with the new setting or abort the
measurement.
The system will not allow a Detector/Disc. Input Frequency of less
than 5 MHz. If you are using an external Phase Detector at a
frequency < 5MHz, enter the fundamental frequency as the carrier
frequency and 5 MHz or the Detector/Disc, frequency. (Note that
when measuring frequency < 5 MHz you must filter out the
fundamental carrier external to the HP 11848A.)
VCO Tune Port Parameters
VCO Tuning Constant is the resulting frequency deviation of the Voltage
Controlled Oscillator (VCO) for a 1 volt change at its tune port. This entry
must be accurate to within ± 5% when the HP 3048A is not going to
measure the VCO Tuning Constant. The entry must be accurate to within
a factor of 2 when the VCO Tuning Constant is going to be measured by
the HP 3048A.
2-16 Instr. Params: $N w/PLL