2.
Access the 2 Oscillator Comparison display from the main software
level by pressing
Access Graph
Other Keys Cmputd Outputs
2 Osc. Compar.
NOTE
The system expects to find the results for the 2 Oscillator
Comparison stored in the defined mass storage.
3.
Enter the title for each of the devices into the Oscillator A and B rows.
(You are limited to 35 characters in the title.) We suggest using device
related nomenclatures and serial numbers as a method for titling the
devices.
4.
Press the
A
VS.
B
File softkey to select the result file for the devices
measured. You are required to select the proper files to be compared
so that device (A), in (A vs. B) is the known device and (B) is the
unknown device.
5.
Press the
Freeze Files
softkey to return to the Oscillator A entry and
freeze the comparison files.
softkey to start the computation.
6. Press the
Solve For B
7.
Once the computation is complete, the system stores the results for the
unknown device in the Result Directory. *ONE OSC* is added to the
file title to indicate that the data was computed, and only represents
the noise level of one device.
8. To recall the results of the DUT, access the Results Directory by
pressing
DONE Result Files
9. To load the Result File, position the cursor at the appropriate title in
the directory and press Load File
10.
To display the noise graph, press
DONE
Other Keys
Plot Data
2 Osc. Compar. 4-25