Agilent Technologies HP 8642A/B Specifiche Pagina 113

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 453
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 112
Parameter Entry Descriptions
Use the Current Detector Constant
When measuring similar devices the "Use the Current Detector" method
can be used to save calibration time. The system assumes that the
frequency and amplitude are identical to the previous measurement in
which the Phase Detector Constant was established. In most cases, the
accuracy uncertainty is negligible as long as the power levels are very
close (±0.3 dB) to the levels present when the Phase Detector Constant
was established.
User entry of the Phase Detector Constant (within the range of Acceptable
Values displayed) is possible when the sensitivity of the detector being
used is known. Entering the Phase Detector Constant can be helpful when
the system is unable to measure the Phase Detector Constant. Keep in
mind that the accuracy of the measurement is dependent on the accuracy
of the Phase Detector Constant. The HP 3048A verifies the accuracy of the
entered Phase Detector Constant when it is configured to verify the Phase
Lock Loop Suppression. An Accuracy Specification Degradation will be
reported if the entered Phase Detector Constant is in error.
The sensitivity of the phase detector is a function of the input level to
the phase detector. You can estimate the Phase Detector Constant directly
from the input power level at the R port of the phase detector; providing
the input level required for driving the L port of the phase detector has
been met. A power meter or a spectrum analyzer can be used to measure
the input level at the R port. Compare the measured R port level to graph
below to estimate the Phase Detector Constant.
2-44 Calibr Process: $N w/PLL
Vedere la pagina 112
1 2 ... 108 109 110 111 112 113 114 115 116 117 118 ... 452 453

Commenti su questo manuale

Nessun commento