Description
The Default Table Values
The default FFT Segment Table provides an optimum set of values for
most noise measurements. The default Segment Table values should be
used the first time a new measurement is made. This guarantees that any
measurement problems are not due to the defined segment table.
Storing the Table
Segment Table data is stored whenever a Test File is created. When a Test
File is loaded, the Segment Table data is set to the values stored in the
Test File.
Modifying the Table
Changes made to the Segment Table do not affect measurement accuracy,
but changes can cause gaps in the Results Graph if the segments do not
cover a continuous frequency range. Also, changes can result in longer
then necessary measurement periods.
Any time the Segment Table is changed and Repeat Msrmnt is selected the
entire frequency offset range of interest should be remeasured, as existing
segment data may be overwritten by new data.
Considerations for Modifying the Segment Table
There is a basic trade-off between the resolution of a measurement and
the time the measurement takes. To increase measurement resolution, fre-
quency, or amplitude, you must be willing to accept longer measurement
times.
It may be possible, however, to minimize the increase in measure-
ment time and still get the increased measurement resolution you need
over a specific portion of the graph by decreasing the resolution for other
portions of the graph.
5-14 FFT Segmnts