Table 1-2. HP 3048A System Option Specifications (1 of 9)
HP 3048A
Option
Specifications
The HP 3048A can be ordered with any of several optional
signal generators as a reference source for phase noise meas-
urements. The following specifications address system opera-
tion with these signal generators. The data that follows is in
addition to that given previously under the heading of
HP 3048A System Specifications. Refer to the data sheet for
each signal generator for more complete information on each
model.
OPTIONS 001 OR 002: ADDING THE HP 8662A
OR 8663A SIGNAL GENERATOR
The following data applies only if either the HP 8662A Opt.
003 or 8663A Opt. 003 is used as the reference source to
demodulate the test signal.
Frequency
Range: 100 kHz to 1280 MHz (to 2560 MHz with HP 8663A).
1
Resolution: 0.1 Hz, 0.2 Hz: 640 to 1280 MHz, 0.4 Hz above
1280 MHz.
Accuracy and Stability (internal 10 MHz quartz oscillator):
Aging rate <5 X 10~
10
/day after 10-day warm-up (typically
24 hrs in normal operating environment).
EFC:
Provides a drift tracking range of ±10
-8
with no degrada-
tion of phase noise or spurious.
Measurements <5 MHz require external phase detector.
1-16 General Information