The Loop Suppression Verification
Introduction
Verification of the Phase Lock Loop Suppression insures that the Phase
Lock Loop Bandwidth (PLL BW), the Phase Detector Constant, and the
VCO Tuning Constant are all accurate for the measurement. (Note that the
HP 3048A always corrects for the loop suppression whether verification is
selected or not.)
Description
The HP 3048A verifies Loop Suppression before beginning the measure-
ment by closing the phase lock loop and injecting noise into the loop. The
response of the loop is measured and recorded. The loop is then opened,
and a measurement is made of the open loop response. The suppression of
the PLL is the difference between the closed loop response and open loop
response to the injected noise. The measured Loop Suppression curve is
smoothed and then four points are selected along the curve including the
peak and the 8 dB point. The measured curve is then compared to the
theoretical curve at the four selected points. Figure 2-16 shows an example
of a Loop Suppression curve measured by the HP 3048A. (For informa-
tion about displaying and interpreting the curve, and using the 11848A
Control softkey for troubleshooting this measurement, refer to Test Mode
in Chapter 5 Special Functions.
The amount of adjustment required to fit the theoretical curve to the mea-
sured curve is recorded as the PLL GAIN CHANGE on the Suppression
Graph and the frequency of the Assumed Pole. The Assumed Pole is a
software controlled term used to fit the Adjusted Theoretical loop response
to the smoothed measured response of the phase lock loop. The MAX ER-
ROR is the final amount of error existing between the adjusted theoretical
curve and the measured curve of the phase lock loop suppression. If the
PLL GAIN CHANGE or MAX ERROR is greater than 1 dB, the System
2-54 Calibr Process: $N w/PLL
rev.05SEP89