The sensitivity of the HP 11848A front panel meter depends on the gain
and attenuation settings within the phase lock circuitry. When the system
is phase locked, 0 V (center scale) on the meter indicates that the sources
are tuned to exactly the same frequency. The HP 3048A detects an out of
lock indication when the meter moves off zero more than ±2.5 divisions.
The full scale meter voltage is determined from the sum of the Center
Voltage of the VCO Tuning Curve (CV) and the Voltage Tuning Range of
VCO (VTR) entered in the Instr Parameter Menu and the configuration of
the HP 11848A's internal attenuator. The following table shows the four
full scale meter voltages possible.
Full Scale
Meter
Voltage
1
2
5
10
Center Voltage (CV)
Voltage Tuning
Range (VTR)
0.2V < CV +VTR < 1.2V
1.2V < CV + VTR < 2.4V
2.4V < CV + VTR < 6V
6V < CV +VTR < 12V
Spurs on the input signals to the phase detector must be less than
-26 dBc at offsets greater than the PLL BW. At offsets within the PtX
BW, spurs may increase at 20 dB per decade starting at —26 dBc at the
PLL BW.
Harmonics on the input signals to the phase detector must be less than
20 dBc or harmonics of a square wave.
Odd harmonics on the beatnote out of the phase detector must be less
than —10 dBc and even harmonics must be less than —30 dBc.
Source Return Loss less than 9.5 dB (2:1 SWR) may degrade the
performance of the HP 3048A system.
Calibr Process: $N w/PLL 2-49