Table 8-4. Warning Messages
SYSTEM UNINITIALIZED. PROPER OPERATION UNLIKELY... The system is unable to
find the default files. Make sure the System Data Disc is in the correct mass storage location.
Action:
Press System Preset to initialize and load the required files. Make sure the System Data Disc
is in the correct Mass Storage location.
(instrument) is not responding...
Action:
Check that the correct HP-IB addresses have been entered into the System Configuration
Table.
Also check that the instrument is connected via HP-IB, and that
its power is ON.
(instrument) is an unsupported (source, analyzer, etc..) The system does not support
this instrument, and the instrument is configured in the System Configuration Table.
rev.05SEP89 Warning Messages 8-28.1