Use the Current Discriminator Constant
Introduction
This calibration method configures the system to use the Discriminator
Constant currently in memory. This eliminates the time spent calibrating
the Discriminator Constant. The Discriminator Constant in memory is the
value that was determined by the last measurement made (unless a stored
Test File has been loaded). The Discriminator Constant value appears in
the Calibration Process display.
The Discriminator Constant is valid for repeat measurements of similar
devices as long as the delay and the input power levels remain the same
(within ±0.3 dB).
Parameter Entry Descriptions
Entering a Discriminator Constant Value
The Discriminator Constant's value may be changed by keying in the
desired value. The Acceptable Limits for the Discriminator Constant are
shown on the display. (Keep in mind that absolute measurement accuracy
depends on the accuracy of the Discriminator Constant.)
Calibr Process: $N w/FM Discriminator 2-77