Table 8-5. Status Messages (9 of 11)
SYSTEM OPERATING IN PULSE MODE. ESTABLISH QUADRATURE VIA AUX.
MONITOR PORT... The PULSED Carrier Type was selected and the system cannot verify
quadrature or establish quadrature automatically.
Action:
Use an oscilloscope connected to the HP 11848A Aux Monitor Port to achieve
quadrature (dc). Refer to Carrier Type in Chapter 5, Special Functions for Information on
establishing quadrature with a Pulsed Measurement.
Taking (RF_Model) sweep... This message indicates that the System is performing a
measurement
measurement.
Taking HP 3561A sweep... This message indicates that the System is performing a
measurement.
TEST PATH: REFERENCE PATH... Indicates which HP 11848A internal path is being used
for a reference measurement in a transfer function measurement.
The FM Deviation on the () will be changed from (X) to (Y). If this is unacceptable,
press 'Abort'... During auto zerobeat of sources, the reference source frequency changed
invalidating the FM Deviation setting specified for it.
Action:
Press
Proceed or Abort
The LNA is currently (in/out) This may be changed via the Toggle LNA' key... When
PULSED is selected for the 'Carrier Type' in the Special Function Menu, a 'Toggle LNA' key
will appear in some screens to allow the user to select between having the LNA in or out. If
the system has decided to switch OUT the LNA and you are concerned about measurement
sensitivity, you may wish to further evaluate the decision. Connect an oscilloscope to the
AUX Monitor Port on the HP 11848A. Observe the LNA output and toggle the LNA IN. If the
peak output level exceeds —0.5V or +1V, the noise may be clipped or compressed so the
measurements must be made with the LNA OUT. If the noise is not significantly clipped, the
measurement may be made with the LNA IN.
The LNA of the HP11848A has been bypassed. This will result in a degradation of the
system noise floor... This is a status message indicating a high input level into HP 11848A. If
you wish to further evaluate the LNA decision, abort the measurement and access the Special
Functions display. Select the PULSED Carrier Type and then initiate a Repeat Measurement.
When the LNA decision is displayed again, connect an oscilloscope to the AUX Monitor port or
the HP 11848A. Observe the LNA output and toggle the LNA IN. If the peak output level exceeds
—0.5V or +1V, the noise may be clipped or compressed so the measurements must be made
with the LNA OUT. If the noise is not significantly clipped, the measurement may be made with
the LNA IN.
The plotted data was derived from Two-oscillator or Three-oscillator comparison... This
message is shown as the parameter summary for a result that was computed, not measured.
Status Messages 8-37