Description of the Test Files
Test Files Shipped with the HP 3048A
When the HP 3048A is shipped, it includes several predefined Test Files.
The Test Files included in the Data Disc are described below.
The Default file sets up the HP 3048A to measure the HP 11848A's Internal
10 MHz Sources (10 MHz A vs. 10 MHz B). This measurement provides
a quick verification of the HP 3048A's ability to complete a Phase Noise
measurement. The results for this measurement may be compared with
the Default Result File also stored on the Data Disc.
NOTE
Any Test File labeled "Default" at the location "Test Param-
eter Files" specified in the Mass Storage Media table will be
loaded automatically every time the HP 3048A Software is
loaded or the
System Preset
softkey is pressed.
Files Labeled HEWLETT-PACKARD FACTORY.... contain predefined Test Files
used for performing the HP 3048A Performance Tests. (Please refer to the
HP 3048A Calibration Manual for details on when and how to perform
these tests.)
Files Labeled HP EXAMPLE.... contain predefined test parameters. The pre-
defined parameters are typical values for measuring the devices indicated
by the Test File labels. These files are provided to aid you in setting up
your own measurements. (Details for applying or modifying these files to
meet your needs are provided in the HP 3048A Operating Manual.)
Files Labeled
HP
TEST.... contain predefined Performance Tests required to
verify operation of the HP 3048A System and its supported options. For
example, HP TEST HP 11729C/8662/63 vs HP 11729C/8662/63 is the Noise
2-122 Test Files
rev.05SEP89