Consideration for Result Files
Copying the Result Files using copy routines provided in the Basic
Operating System will not update the Result File Directory and may cause
loss of data. Copy Result Files from one disc to another by loading the
Result File into the System and then Storing the Result File on to the other
disc.
When loading a result file, be certain you have specified the correct graph
type for the file being loaded. For instance, if you have selected the graph
type for a Phase Noise Using PLL measurement, and then you load a result
file for an AM Noise measurement, an error message is displayed to alert
you to the condition.
When a Result File is loaded, the Carrier Frequency parameter entered into
the Instrument Parameter display is not changed. If you plan to compute
Sigma vs. Tau for the loaded Result File, first verify that the defined Carrier
Frequency is appropriate for the loaded file. An incorrect Carrier Frequency
entry will produce invalid Sigma vs. Tau computations. (Carrier Frequency
is the only measurement parameter that is used for the Sigma vs. Tau
computation).
The "Resu_" numbers represent the media file name and the order in which
the Result Files were stored.
rev.05SEP89
Result Files 3-15