Table 6-7. Supported RF Analyzers
Model
HP 3585A/B
HP 8566A/B
HP 8567A
HP 8568A/B
HP 71000
Frequency Range
20 Hz to 40 MHz
100 Hz to 22 GHz
1 kHz to 1500 MHz
100 Hz to 1500 MHz
Varies with model
Other Considerations
Allows complete generation of calibration
files,
Synthesized LO, Tracking Generator
Must verify calibration data by performing
the Noise Flatness Performance Test.
Synthesized LO. Although this analyzer
cannot generate New CALDATAHI
> 100 kHz, the Cal Data is verified and/or
corrected by performing the Noise Flatness
Test.
Must verify calibration data by performing
the Noise Flatness Performance Test.
Synthesized LO. Although this analyzer
cannot generate New CALDATAHt
> 100 kHz, the Cal Data is verified and/or
corrected by performing the Noise Flatness
Test.
Must verify calibration data by performing
the Noise Flatness Performance Test.
Synthesized LO. Although this analyzer
cannot generate New CALDATAHI
> 100 kHz, the Cal Data is verified and/or
corrected by performing the Noise Flatness
Test.
Must verify calibration data by performing
the Noise Flatness Performance Test.
Synthesized LO. Although this analyzer
cannot generate New CALDATAHI
> 100 kHz, the Cal Data is verified and/or
corrected by performing the Noise Flatness
Test. Although a tracking generator is
available, it is not supported by HP 3048A
software for generation of cafibration files.
6-24 RF Analyzers