2- 62 Agilent B2900 SCPI Command Reference, Edition 2
Subsystem Command Summary
Using Advanced Functions
Table 2-22 STATus Subsystem
Command Summary Reset setting
:STAT:MEAS:COND?
:STAT:OPER:COND?
:STAT:QUES:COND?
Returns the value of the measurement,
operation, or questionable status
condition register.
:STAT:MEAS:ENAB mask
:STAT:OPER:ENAB mask
:STAT:QUES:ENAB mask
:STAT:MEAS:ENAB?
:STAT:OPER:ENAB?
:STAT:QUES:ENAB?
Sets the measurement, operation, or
questionable status enable register.
mask=0 to 65535 (decimal)
0
:STAT:MEAS[:EVEN]?
:STAT:OPER[:EVEN]?
:STAT:QUES[:EVEN]?
Returns the value of the measurement,
operation, or questionable status event
register.
:STAT:MEAS:NTR filter
:STAT:OPER:NTR filter
:STAT:QUES:NTR filter
:STAT:MEAS:NTR?
:STAT:OPER:NTR?
:STAT:QUES:NTR?
Sets the negative transition filter in the
measurement, operation, or questionable
status register. If you set a bit of the filter,
a 1-to-0 transition of its register bit sets
the corresponding bit of the event
register.
filter=0 to 65535 (decimal)
0
:STAT:MEAS:PTR filter
:STAT:OPER:PTR filter
:STAT:QUES:PTR filter
:STAT:MEAS:PTR?
:STAT:OPER:PTR?
:STAT:QUES:PTR?
Sets the positive transition filter in the
measurement, operation, or questionable
status register. If you set a bit of the filter,
a 0-to-1 transition of its register bit sets
the corresponding bit of the event
register.
filter=0 to 65535 (decimal)
32767
:STAT:PRES
Sets all defined bits in the status system’s
PTR registers and clears the all bits in the
NTR and Enable registers. The registers
are returned to the default condition.
Vedere la pagina 101