Agilent B2900 SCPI Command Reference, Edition 2 2- 23
Subsystem Command Summary
Setting Source/Measure Unit
:OUTP[c]:OFF:MODE mode
:OUTP[c]:OFF:MODE?
Selects the source condition after output
off.
mode=ZERO|HIZ|NORMal
NORM
:OUTP[c]:ON:AUTO mode
:OUTP[c]:ON:AUTO?
Enables or disables the automatic output
on function.
mode=1|ON|0|OFF
ON
:OUTP[c]:PROT[:STAT] mode
:OUTP[c]:PROT[:STAT]?
Enables or disables the over voltage/
current protection.
mode=1|ON|0|OFF
ON
:OUTP[c]:REC index
Recalls the channel setup.
index=0 or 1 to recall the channel setup 0
or 1 saved by the :OUTP:SAVE
command.
:OUTP[c]:SAVE index
Saves the channel setup. The setup can be
recalled by the :OUTP:REC command.
index=0 or 1 to memorize the present
channel setup as the channel setup 0 or 1.
:OUTP[c][:STAT] mode
:OUTP[c][:STAT]?
Enables or disables the source output.
mode=1|ON|0|OFF
OFF
Command Summary Reset setting
Vedere la pagina 62