Agilent B2900 SCPI Command Reference, Edition 2 2- 11
Subsystem Command Summary
Setting Source/Measure Unit
[:SOUR[c]]:LIST:CURR:APP append_
list
[:SOUR[c]]:LIST:VOLT:APP append_
list
Adds the source output current or voltage
data to the end of the list set by the
[:SOUR[c]]:LIST:<CURR|VOLT>
command, to which some data might be
appended to by this command. Total
number of data in the list must be
2500.
append_list: List of the output current or
voltage data. Multiple data can be set to
append_list. Each data must be separated
by a comma.
[:SOUR[c]]:LIST:CURR:POIN?
[:SOUR[c]]:LIST:VOLT:POIN?
Returns the number of data in the list set
by the [:SOUR[c]]:LIST:<CURR|VOLT>
command, to which some data might be
appended to by the
[:SOUR[c]]:LIST:<CURR|VOLT>:APP
command.
[:SOUR[c]]:LIST:CURR:STAR start
[:SOUR[c]]:LIST:VOLT:STAR start
[:SOUR[c]]:LIST:CURR:STAR?
[:SOUR[c]]:LIST:VOLT:STAR?
Specifies the list sweep start point by
using the index of the list.
start: Index of the list. 1 to 2500. start=1
indicates the first data in the list (top of
the list). start=0 or the value greater than
2500 causes an error.
1
[:SOUR[c]]:PULS:DEL delay
[:SOUR[c]]:PULS:DEL? [DEFault|MI
Nimum|MAXimum]
Sets the pulse delay time for the specified
channel. The pulse delay time is the time
from starting the pulse base output to
starting the pulse level transition (or to
starting the pulse peak output). See
Figures
2-4 and 2-2.
delay=DEFault|MINimum|MAXimum|
0.0 to 99999.9 seconds
0
Command Summary Reset setting