5- 10 Agilent B2900 SCPI Command Reference, Edition 2
Error Messages
Device-Dependent Error
Device-Dependent Error
-3XX errors indicate that Agilent B2900 has detected an error that is not a
command error, a query error, or an execution error; some device operations did not
properly complete, possibly due to an abnormal hardware or firmware condition.
These codes are also used for self-test response errors.
Error -300 Device-specific error
Generic device-dependent error for B2900 that cannot be determined more
specifically.
Error -310 System error
Some error, termed “system error” by B2900, has occurred.
Error -311 Memory error
An error was detected in B2900’s memory.
Error -313 Calibration memory lost; Calibration data has been lost,
Calibration data is initialized; Channel n
Calibration memory lost; Nonvolatile data saved by the
*CAL? command has been lost; Channel n
Non-volatile data related to the *CAL? command has been lost.
Error -315 Configuration memory lost
Non-volatile configuration data saved by B2900 has been lost.
Error -321 Out of memory
Too many data was sent at a time.
Error -350 Queue overflow
This code is entered into the queue instead of the code that caused the error. This
code indicates that there is no room in the queue and an error occurred but was not
recorded.