Agilent B2900 SCPI Command Reference, Edition 2 4- 103
Subsystem Commands
[:SOURce]:FUNCtion:MODE
:SOUR:DIG:INT2:TOUT:POS?
[:SOURce]:FUNCtion:MODE
Selects the source output mode of the specified channel.
Syntax [:SOURce[c]]:FUNCtion:MODE mode
[:SOURce[c]]:FUNCtion:MODE?
Parameter mode Source output mode. CURRent|VOLTage (default). Parameter
data type is CPD.
mode=CURR sets the specified channel to the current source. Voltage compliance
for the current source is set by the :SENS:VOLT:PROT[:LEV] command.
mode=VOLT sets the specified channel to the voltage source. Current compliance
for the voltage source is set by the :SENS:CURR:PROT[:LEV] command.
See “:SENSe:<CURRent[:DC]|VOLTage[:DC]>:PROTection[:LEVel]” on page
4-76.
Query response mode <newline>
mode returns CURR or VOLT. Response data type is CRD.
Example :FUNC:MODE CURR
:SOUR2:FUNC:MODE?
[:SOURce]:FUNCtion[:SHAPe]
Selects the source output shape of the specified channel.
Syntax [:SOURce[c]]:FUNCtion[:SHAPe] shape
[:SOURce[c]]:FUNCtion[:SHAPe]?
Parameter shape Source output shape. PULSe|DC (default). Parameter data type
is CPD.
shape=DC sets the specified channel to DC (constant) output.
shape=PULS sets the specified channel to pulsed output.
Query response shape <newline>