
Agilent B2900 SCPI Command Reference, Edition 2 4- 115
Subsystem Commands
:STATus:<MEASurement|OPERation|QUEStionable>:CONDition?
STATus Subsystem
:STATus:<MEASurement|OPERation|QUEStionable>:
CONDition?
Returns the value of the measurement, operation, or questionable status condition
register. See
Table 4-4 to 4-6 for the bit definitions. The register setting is not
changed by this command.
Syntax :STATus:<MEASurement|OPERation|QUEStionable>:CONDition?
For <MEASurement|OPERation|QUEStionable>, specify MEASurement for the
measurement status condition register, OPERation for the operation status condition
register, or QUEStionable for the questionable status condition register.
Query response value <newline>
value returns the value of the specified register. It is the sum of the binary-weighted
values for the set bits. Response data type is NR1 (decimal) or NDN (binary, octal,
or hexadecimal) selected by the
:FORMat:SREGister command.
Example :STAT:MEAS:COND?
:STAT:OPER:COND?
:STAT:QUES:COND?
Table 4-4 Questionable Status Condition Register Bit Definitions
bit
decimal
value
description definition
0 1
Voltage Summary
Over voltage in channel 1, and/or 2.
1 2
Current Summary
Over current in channel 1, and/or 2.
2 4 Ch1 Output Protection Output relay of the specified channel is opened
by the automatic output off at compliance
function.
3 8 Ch2 Output Protection
4 16
Temperature Summary
Over temperature in channel 1, and/or 2.
5 to 7 Not used 0 is returned.
Commenti su questo manuale