
2- 16 Agilent B2900 SCPI Command Reference, Edition 2
Subsystem Command Summary
Setting Source/Measure Unit
:SENS[c]:CURR[:DC]:NPLC:AUTO mod
e
:SENS[c]:RES:NPLC:AUTO mode
:SENS[c]:VOLT[:DC]:NPLC:AUTO mod
e
:SENS[c]:CURR[:DC]:NPLC:AUTO?
:SENS[c]:RES:NPLC:AUTO?
:SENS[c]:VOLT[:DC]:NPLC:AUTO?
Enables or disables the automatic NPLC
function.
mode=1|ON|0|OFF
The automatic NPLC on/off works with
the automatic aperture on/off set by the
:SENS:<CURR|RES|VOLT>:APER:AU
TO command. So the last command
setting is effective for both functions.
OFF
:SENS[c]:CURR[:DC]:PROT[:LEV] co
mp
:SENS[c]:VOLT[:DC]:PROT[:LEV] co
mp
:SENS[c]:CURR[:DC]:PROT[:LEV]? [
DEFault|MINimum|MAXimum]
:SENS[c]:VOLT[:DC]:PROT[:LEV]? [
DEFault|MINimum|MAXimum]
Sets the compliance value of the specified
channel.
comp=MINimum|MAXimum|DEFault|
minimum to maximum measurement
value, in A or V. See
“Measurement
Ranges” on page 2-27.
1.00E-04 A
for current
compliance,
2.0 V for
voltage
compliance
:SENS[c]:CURR[:DC]:PROT:TRIP?
:SENS[c]:VOLT[:DC]:PROT:TRIP?
Returns the compliance status of the
specified channel
Response is 1 or 0 that indicates the
channel is in the compliance state or not.
:SENS[c]:CURR[:DC]:RANG:AUTO mod
e
:SENS[c]:RES:RANG:AUTO mode
:SENS[c]:VOLT[:DC]:RANG:AUTO mod
e
:SENS[c]:CURR[:DC]:RANG:AUTO?
:SENS[c]:RES:RANG:AUTO?
:SENS[c]:VOLT[:DC]:RANG:AUTO?
Enables or disables the automatic ranging
function of the specified measurement
channel.
mode=1|ON|0|OFF
ON
Command Summary Reset setting
Commenti su questo manuale