Agilent B2900 SCPI Command Reference, Edition 2 4- 25
Subsystem Commands
:DISPlay[:WINDow]:DATA?
:DISPlay:VIEW?
Parameter mode SINGle1|SINGle2|DUAL|GRAPh|ROLL. Parameter data type
is CPD.
mode=SINGle1 sets the channel 1 display mode. Default setting for 1-channel
models.
mode=SINGle2 sets the channel 2 display mode. Only on 2-channel models.
mode=DUAL sets channel 1, and channel 2 display mode. Only on 2-channel
models. Default setting.
mode=GRAPh sets the graph display mode for the sweep measurement results.
mode=ROLL sets the roll display mode for the time domain measurement results.
Query response mode <newline>
mode returns SING1, SING2, DUAL, GRAP, or ROLL. Response data type is CRD.
Example :DISP:VIEW GRAP
:DISP:VIEW?
:DISPlay[:WINDow]:DATA?
Returns the data displayed on the front panel display.
Syntax :DISPlay[:WINDow[d]]:DATA?
Query response response <newline>
response returns the measured values, source output value, and sense limit value
displayed on the front panel display, as shown below. Each data is separated by a
comma. response returns ----------- (hyphen) for the empty data. Response data type
is SRD.
meas_value1,meas_value2,source_output,sense_limit
Characters and are converted to u and ohm, respectively.
For the SING1, SING2, or DUAL display mode, the :DISP:DATA? command
returns the data displayed on the upper display area.
For the SING1 or SING2 display mode, the :DISP:WIND2:DATA? command
returns “-----------,-----------,-----------,-----------”.
Vedere la pagina 146