4- 4 Agilent B2900 SCPI Command Reference, Edition 2
Subsystem Commands
:CALCulate:CLIMits:CLEar[:IMMediate]
Parameter time value (+1E-5 to 60 seconds)|MINimum|MAXimum|DEFault
(default is +1E-4). Parameter data type is NRf+. Query does
not support time=value. If you specify the value less than MIN
or greater than MAX, time is automatically set to MIN or MAX.
Query response time <newline>
time returns the present setting of delay time for the automatic clear. If a parameter
is specified, time returns the value assigned to DEF, MIN, or MAX. Response data
type is NR3.
Example :CALC:CLIM:CLE:AUTO:DEL 1E-3
:CALC2:CLIM:CLE:AUTO:DEL?
:CALCulate:CLIMits:CLEar[:IMMediate]
Clears the composite limit test results and ports (GPIO lines) immediately.
Syntax :CALCulate[c]:CLIMits:CLEar[:IMMediate]
Example :CALC:CLIM:CLE:IMM
:CALC2:CLIM:CLE
:CALCulate:CLIMits:<FAIL|PASS>:DIGital[:DATA]
Defines a fail/pass pattern that appears near the end of the flowcharts shown in
Figures
2-10 and 2-11. This is a bit pattern used to indicate the composite limit test
result (fail or pass). It must be entered in the format set by the :FORMat:DIGital
command.
Syntax :CALCulate[c]:CLIMits:<FAIL|PASS>:DIGital[:DATA] bit_pattern
:CALCulate[c]:CLIMits:<FAIL|PASS>:DIGital[:DATA]?
For <FAIL|PASS>, specify PASS for a pass pattern or FAIL for a fail pattern.
Parameter bit_pattern 0 (default setting in decimal expression) to 16383. Parameter
data type is NR1 or NDN.
Query response bit_pattern <newline>
bit_pattern returns the fail/pass bit pattern in the format specified by the
:FORMat:DIGital command. Response data type is NR1 or NDN.
Commenti su questo manuale