5- 14 Agilent B2900 SCPI Command Reference, Edition 2
Error Messages
B2900 Specific Error
Error 214 Display must be enabled
Display is currently disabled. Set remote display on.
Error 215 Remote sensing must be on
Remote sensing must be on to perform the requested operation.
Error 216 Auto resistance measurement must be off
Automatic resistance measurement must be off to perform the requested operation.
Error 301 Emergency; Overvoltage status detected; Channel n
Overvoltage status was detected in channel n. All channel output is changed to 0 V
and the all output switch is opened. Execute the *TST? command.
Error 302 Emergency; Overcurrent status(245 V) detected; Channel n
overcurrent status (245 V) was detected in channel n. All channel output is changed
to 0 V and the all output switch is opened. Execute the *TST? command.
Error 303 Emergency; Overcurrent status(35 V) detected; Channel n
overcurrent status (35 V) was detected in channel n. All channel output is changed
to 0 V and the all output switch is opened. Execute the *TST? command.
Error 304 Emergency; Over range current status detected; Channel n
Over range current status was detected in channel n. All channel output is changed
to 0 V and the all output switch is opened. Execute the *TST? command.
Error 305 Emergency; High temperature1 status detected; Channel n
High temperature 1 status was detected in channel n. All channel output is changed
to 0 V and the all output switch is opened. Execute the *TST? command.
Error 306 Emergency; High temperature2 status detected; Channel n
High temperature 2 status was detected in channel n. All channel output is changed
to 0 V and the all output switch is opened. Execute the *TST? command.
Error 307 Emergency; High temperature3 status detected; Channel n
High temperature 3 status was detected in channel n. All channel output is changed
to 0 V and the all output switch is opened. Execute the *TST? command.