4- 88 Agilent B2900 SCPI Command Reference, Edition 2
Subsystem Commands
:SENSe:TOUTput:SIGNal
mode=1 or ON enables offset-compensated resistance measurement.
Query response mode <newline>
mode is 0 or 1, and indicates that offset-compensated resistance measurement is off
or on, respectively. Response data type is NR1.
Example :SENS:RES:OCOM 1
:SENS2:RES:OCOM?
:SENSe:TOUTput:SIGNal
Selects the trigger output for the status change between the trigger layer and the
acquire device action. Multiple trigger output ports can be set.
Syntax :SENSe[c]:TOUTput:SIGNal output{,output}
:SENSe[c]:TOUTput:SIGNal?
Parameter output Trigger output port.
EXT1
(default)
|EXT2|EXT3|EXT4|EXT5|
EXT6|EXT7|EXT8|EXT9|EXT10|EXT11|EXT12|EXT13|EXT14|
LAN|INT1|INT2
. Parameter data type is CPD.
output=INT1 or INT2 selects the internal bus 1 or 2, respectively.
output=LAN selects a LAN port.
output=EXTn selects the GPIO pin n, which is an output port of the Digital I/O
D-sub connector on the rear panel. n=1 to 14.
Query response response <newline>
response returns the present setting, INT1, INT2, LAN, or EXT1 through EXT14.
Response data type is CRD. Multiple responses are separated by a comma.
Example :SENS:TOUT:SIGN EXT3
:SENS2:TOUT:SIGN?
:SENSe:TOUTput[:STATe]
Enables or disables the trigger output for the status change between the trigger layer
and the acquire device action.
Syntax :SENSe[c]:TOUTput[:STATe] mode