Agilent B2900 SCPI Command Reference, Edition 2 4- 3
Subsystem Commands
:CALCulate:CLIMits:CLEar:AUTO
CALCulate Subsystem
For the numeric suffixes [c], [m], and [n], see “Numeric Suffix” on page 1-8.
:CALCulate:CLIMits:CLEar:AUTO
Enables or disables the automatic clear function of the composite limit test.
Syntax :CALCulate[c]:CLIMits:CLEar:AUTO mode
:CALCulate[c]:CLIMits:CLEar:AUTO?
Parameter mode 0|OFF|1|ON (default). Parameter data type is boolean.
mode=1 or ON enables the automatic clear function which clears the composite
limit test results and ports (GPIO lines) automatically with each :INITiate
command. See
“:INITiate[:IMMediate]<:ACQuire|:TRANsient|[:ALL]>” on page
4-157.
mode=0 or OFF disables the automatic clear function. The composite limit test
results and ports (GPIO lines) must be cleared manually before the next composite
limit test is started. Execute the
:CALCulate:CLIMits:CLEar[:IMMediate]
command to clear them immediately.
Query response mode <newline>
mode is 0 or 1, and indicates that the automatic clear function is off or on,
respectively. Response data type is NR1.
Example :CALC:CLIM:CLE:AUTO 1
:CALC2:CLIM:CLE:AUTO?
:CALCulate:CLIMits:CLEar:AUTO:DELay
Sets the delay time for the automatic clear of the composite limit test. See
“:CALCulate:CLIMits:CLEar:AUTO” on page 4-3. The delay time is defined as the
time before the automatic clear is performed after the measurement is completed.
Syntax :CALCulate[c]:CLIMits:CLEar:AUTO:DELay time
:CALCulate[c]:CLIMits:CLEar:AUTO:DELay? [time]